Devices available without vacuum pump (types 5301, 5501, 5510) or with internal vacuum pump (types 5301P, 5501P).
Efficient particle counters with specifications according to ISO 21501-4 and JIS B9921 standards.
It can be easily integrated into the building automation and cleanroom monitoring system via Ethernet or (as an optional extra) wireless 802.11 b/g, RS485 or RS232 connection.
The devices display and save data for 6 user-selected particle size channels and flow rates. In addition, the optional temperature/humidity sensor (0 - 50 ºC ± 0.ºC, 15-90% ±2% relative humidity) measures and displays the temperature and relative humidity. They can be controlled and monitored remotely via a web browser. With a colored, 10.9 cm WQVGA (480 x 272) touchscreen, an adjustable, built-in sound-based alarm that counts alarms from 1 to 9,999,999, and a stainless steel exterior.
Count modes:
Particles Plus instruments can store up to 45,000 data records. These records contain comprehensive data for each sample. With 1 sampling per minute, the device can record a full month of data before overwriting the old records. With 1 sampling every 15 minutes, it can store more than 15 months of data records.
They contain industry-leading battery technology, which ensures the longest available operating time for the devices. It is possible to put the devices in partial or full sleep mode for the time between samplings. In sleep mode, the device consumes minimal energy.
If required the devices can also be ordered with custom calibration points.
The manufacturer provides the IMS software for its particle counters free of charge.
It enables the viewing of past and real-time measurements and displays the status of the instrument. The data of the measurements can also be downloaded in a .csv file. |
With the help of the software, the data selected by the user can be directly exported to MS Excel. It also offers the possibility to update the firmware of the instrument and calibrate the environmental sensors (e.g. temperature, relative humidity). |
It provides a quick and easy way to view instrument data. Simply select the data fields and data range for the selected instrument, and the software will retrieve the data and plot it for immediate analysis. |
Specifications: | 5301 | 5501 | 5510 |
---|---|---|---|
Size range: | 0.3 to 25 μm | 0.5 to 25 μm | 0.5 to 25 μm |
Calibrated points: | 0.3 μm, 0.5 μm, 1.0 μm, 3.0 μm, 5.0 μm, 10.0 μm | 0.5 μm, 0.7 μm, 1.0 μm, 3.0 μm, 5.0 μm, 10.0 μm | 0.5 μm, 0.7 μm, 1.0 μm, 3.0 μm, 5.0 μm, 10.0 μm |
Counting efficiency: | 50% at 0.3 μm; 100% for particles > 0.45 μm per JIS | 50% at 0.5 μm; 100% for particles > 0.75 μm per JIS | 50% at 0.5 μm; 100% for particles > 0.75 μm per JIS |
Flow rate: | 2.83 LPM | 2.83 LPM | 28.3 LPM |
Concentration limits: |
15,000,000 particles/ft3 at 5% coincidence loss (424,752.48 particles/m3) |
15,000,000 particles/ft3 at 5% coincidence loss (424,752.48 particles/m3) |
1,500,000 particles/ft3 at 10% coincidence loss (42,475.25 particles/m3) |
Zero count: |
< 1 count / 60 minutes (< 1 particles /6ft3). No fault count subtraction. < 1 count / 60 minutes (< 1 particles /0,17 m3) |
< 1 count / 60 minutes (< 1 particles /6ft3). No fault count subtraction. < 1 count / 60 minutes (< 1 particles /0,17 m3) |
< 1 count / 60 minutes (< 1 particles /6ft3). No fault count subtraction. < 1 count / 60 minutes (< 1 particles /0,17 m3) |
Vacuum requirements: | External vacuum > 38.1 cm of Hg | External vacuum > 38.1 cm of Hg | External vacuum > 38.1 cm of Hg |
Sample time: | 1 second to 99 hours | 1 second to 99 hours | 1 second to 99 hours |
Operating conditions: | 5 °C to 40 °C / 20% to 95% non-condensing | 5 ºC to 40 ºC / 20% to 95% non-condensing | 5 ºC to 40 ºC / 20% to 95% non-condensing |
Storage conditions: | 0 °C to 50 °C / Up to 98% non-condensing | 0 ºC to 50 ºC / Up to 98% non-condensing | 0 ºC to 50 ºC / Up to 98% non-condensing |
Dimensions (L x W x H): | 13.3 cm x 5.8 cm x 22.6 cm | 13.3 cm x 5.8 cm x 22.6 cm | 13.3 cm x 5.8 cm x 22.6 cm |
Weight: | 1.18 kg | 1.18 kg | 1.18 kg |
Specifications: | 5301P | 5501P |
---|---|---|
Size range: | 0.3 to 25 μm | 0.5 to 25 μm |
Calibrated points: | 0.3 μm, 0.5 μm, 1.0 μm, 3.0 μm, 5.0 μm, 10.0 μm | 0.5 μm, 0.7 μm, 1.0 μm, 3.0 μm, 5.0 μm, 10.0 μm |
Counting efficiency: | 50% at 0.3 μm; 100% for particles >0.45 μm per JIS | 50% at 0.5 μm; 100% for particles >0.75 μm per JIS |
Flow rate: | 2.83 LPM | 2.83 LPM |
Concentration limits: |
15,000,000 particles/ft3 at 10% coincidence loss (424,752.48 particles/m3) |
15,000,000 particles/ft3 at 10% coincidence loss (424,752.48 particles/m3) |
Zero count: |
< 1 count / 60 minutes (< 1 particles /6ft3). No fault count subtraction. < 1 count / 60 minutes (< 1 particles /0,17 m3) |
< 1 count / 60 minutes (< 1 particles /6ft3). No fault count subtraction. < 1 count / 60 minutes (< 1 particles /0,17 m3) |
Vacuum requirements: | Internal vacuum pump | Internal vacuum pump |
Sample time: | 1 second to 99 hours | 1 second to 99 hours |
Operating conditions: | 5 ºC to 40 ºC / 20% to 95% non-condensing | 5 ºC to 40 ºC / 20% to 95% non-condensing |
Storage conditions: | 0 ºC to 50 ºC / Up to 98% non-condensing | 0 ºC to 50 ºC / Up to 98% non-condensing |
Dimensions (L x W x H): | 13.3 cm x 5.8 cm x 22.6 cm | 13.3 cm x 5.8 cm x 22.6 cm |
Weight: | 1.18 kg | 1.18 kg |
Particle counters without a vacuum pump or with a built-in vacuum pump that can be integrated into a building automation or facility monitoring system.
More informationsRemote particle counters with a removable battery, so they can also be used as a stand-alone device. Can store up to 45,000 data records.
More informationsHandheld particle counter with removable battery. Can store up to 45,000 data records.
More informationsParticle counters without a vacuum pump or with a built-in vacuum pump that can be integrated into a building automation or facility monitoring system. Can store up to 65,000 data records.